Abstract

Objective: This study aimed to evaluate the accuracy of linear measurements made on CBCT (New Tom VGi) scans with different image magnifications.
 Methods: Forty-two titanium pins were inserted into seven dry sheep mandibles. Lengths of the pins were measured using a digital caliper with 0.01 mm readability, and the mandibles were radiographed using a CBCT unit. After reconstruction of CBCT images, three experienced radiologists measured the length and diameter of titanium pins at100%, 200% and 400% magnifications. Accuracy of measurements was analyzed using descriptive statistics and one-way analysis of variance (ANOVA) P<0.05 was considered statistically significant. Inter observer reliability was calculated using the intra class correlation coefficient (ICC) test.
 Results: The mean differences of linear measurements from the actual lengths of pins were 0.1960 mm, 0.2143 mm and 0.2047 mm at 100%, 200% and 400% magnifications, respectively (P>0.05). The mean differences of linear measurements from the actual diameters of pins were 0.2206 mm, 0.2063 mm and 0.1984 mm at 100%, 200% and 400% magnifications, respectively (P>0.05). Inter observer reliability of pin length measurements was estimated to be0.285, 0.707 and 0.479 at 100%, 200% and 400%magnifications, respectively. Inter observer reliability of pin diameter measurements was 0 .078, 0.469 and 0.587 at 100%, 200% and 400%magnifications, respectively.
 Conclusion: Based on the results, image magnification does not affect the accuracy of linear measurements made on CBCT scans. Inter observer reliability of pin length measurements was good at 200% magnification, while it was below the acceptable range at other magnifications. For measurement of pin diameters, this index was below the acceptable range at all magnifications.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call