Abstract

Objective: To investigate the accuracy of intraoperative electrophysiological studies in detecting incorrectly positioned electrodes in cochlear implant surgery.Study design: A retrospective chart review.Setting: Tertiary referral centre.Patients: In total, 104 consecutive patients with a mean age of 5 years underwent cochlear implant surgery at our centre between January 2012 and December 2013. All patients were implanted with Cochlear Nucleus Freedom implants.Method: A retrospective study to compare intraoperative neural response telemetry (NRT), impedance and electrode position using Stenver’s transorbital plain X-ray view.Results: Intraoperative electrophysiological tests for patients with Cochlear Nucleus Freedom implants showed 97% sensitivity and 100% specificity compared with postoperative X-ray imaging.Conclusion: NRT results for the position of cochlear implants were very accurate when checked by X-ray imaging showing that this technique is sufficient in most cases. Stenver’s plain X-ray view is needed in complicated cases with abnormal NRT testing or difficult electrode insertion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.