Abstract

To assess differences in the accuracy of automatic QT measurement in three subject groups, and to determine the influence of T wave amplitude on these measurements. Standard simultaneous 12 lead electrocardiograms were acquired from 25 patients post myocardial infarction, 25 with arrhythmias, and 25 controls. Because there is not yet a standard automatic method for QT analysis, four different techniques were used. Manual QT measurements were used as the reference. QT was measured in two complexes by each technique in each lead, subject, and group. The differences between reference and automatic QT measurements from the three subject groups were compared independently for the four techniques. The T wave amplitudes for each of the groups were also compared. Variability of the automatic QT measurements, relative to the manual reference, in the cardiac patients was 2.1 times that in the controls (P < 0.005). Mean T wave amplitude was lower (by a factor of two) for the cardiac patients compared with the controls (P < 0.01). No simple relation between T wave amplitude and the difference between automatic and manual QT measurements was found, although the difference was 2.2 times greater for absolute T wave amplitudes of less than 0.25 mV (P < 0.001). Automatic QT measurement techniques are less accurate in cardiac patients than in controls. Measurements from T waves with amplitudes less than 0.25 mV are less reliable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.