Abstract

Aims: To investigate the accuracy of FreeStyle LibreTM flash glucose monitoring (FGM) relevant to plasma glucose (PG) measurements during postprandial rest and different walking conditions in overweight/obese young adults.Methods: Data of 40 overweight/obese participants from two randomized crossover studies were pooled into four trials: (1) sitting (SIT, n = 40); (2) walking continuously for 30 min initiated 20 min before individual postprandial glucose peak (PPGP) (20iP + CONT, n = 40); (3) walking continuously for 30 min initiated at PPGP (iP + CONT, n = 20); and (4) accumulated walking for 30 min initiated 20 min before PPGP (20iP + ACCU, n = 20). Paired FGM and PG were measured 4 h following breakfast.Results: The overall mean absolute relative difference (MARD) between PG and FGM readings was 16.4 ± 8.6% for SIT, 16.2 ± 4.7% for 20iP + CONT, 16.7 ± 12.2% for iP + CONT, and 19.1 ± 6.8% for 20iP + ACCU. The Bland–Altman analysis showed a bias of −1.03 mmol⋅L–1 in SIT, −0.89 mmol⋅L–1 in 20iP + CONT, −0.82 mmol⋅L–1 in iP + CONT, and −1.23 mmol⋅L–1 in 20iP + ACCU. The Clarke error grid analysis showed that 99.6–100% of the values in all trials fell within zones A and B.Conclusion: Although FGM readings underestimated PG, the FGM accuracy was overall clinically acceptable during postprandial rest and walking in overweight/obese young adults.

Highlights

  • Continuous glucose monitoring (CGM), a minimally invasive method of continuously measuring interstitial glucose levels to gain insight into their patterns and trends, has been increasingly adopted as a novel and feasible tool to improve glycemic control in individuals with diabetes and those with poor glucose control

  • Data of 40 overweight/obese participants from two randomized crossover studies were pooled into four trials: (1) sitting (SIT, n = 40); (2) walking continuously for 30 min initiated 20 min before individual postprandial glucose peak (PPGP) (20iP + CONT, n = 40); (3) walking continuously for 30 min initiated at PPGP; and (4) accumulated walking for 30 min initiated 20 min before PPGP (20iP + ACCU, n = 20)

  • Accuracy of Flash Glucose Monitoring much greater insight into glucose levels throughout the day (Klonoff, 2005). As it is widely used in research, CGM allows for opportunities to continuously examine glycemic control and postprandial glucose (PPG), as well as glucose responses to exercise and meals (Macleod et al, 2013)

Read more

Summary

Methods

Paired FGM and PG were measured 4 h following breakfast

Results
Conclusion
INTRODUCTION
MATERIALS AND METHODS
Participants
Study Design and Protocols
RESULTS
DISCUSSION
ETHICS STATEMENT
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.