Abstract

To conduct a meta-analysis on the accuracy of electronic apex locators (EALs) in determining working length in human primary teeth. An electronic literature search was conducted using a combination of the terms 'primary teeth', 'root canal length' and 'electronic apex locators'. Additional eligible articles were identified using a manual search. The full texts of identified articles were screened according to inclusion criteria. Overall, 19 studies met the inclusion criteria and were subjected to further data extraction. Meta-analysis of studies that reported mean lengths and difference in mean lengths between EAL and actual root canal length (ARCL) revealed a significant difference (P=0.015) between the two readings whilst those studies that evaluated intraclass correlation (ICC) suggested a high correlation (P<0.0001) between the EAL and ARCL readings. A subgroup analysis revealed that the presence of root resorption did not affect the accuracy of EALs (P=0.567). Electronic apex locators offer an acceptable level of accuracy in the measurement of root canal length in primary teeth.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.