Abstract

Maximum lateral forces as measured by strain gauges mounted on the analysing rod of a Nesor surveyor for four operators during surveying procedures ranged from 81 to 110 cN (mean 92.4 cN). Maximum lateral forces used when trimming blockout and inlay wax ranged from 159 to 184 cN (mean 169 cN) and 137 to 196 cN (mean 172 cN), respectively. Mean tilting errors produced in the vertical surveying arm by lateral loads of 98 and 196 cN (i.e. weights of 100 and 200 g) were determined to be 0.29 and 0.42° for five Nesor surveyors and 0.21 and 0.36° for five Krupps surveyors, respectively. All surveyors had been in constant use for at least 12 years. Play in the bearings of the horizontal arm due to loads of the same magnitude applied obliquely produced an additional mean error of approximately 0.5° in the Krupps surveyors. Finally, discrepancies of up to approximately ± 0.04 mm were found in 10 undercut gauges in each of the nominal sizes 0.25, 0.50 and 0.75 mm. It is concluded that surveyors and their accessories should be checked and serviced regularly, and possible sources of error pointed out to potential users.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.