Abstract

To evaluate the feasibility and accuracy of cervical spine pedicle screw fixation assisted by X-ray fluoroscopy and CT-based navigation system. 145 cervical pedicle screws were placed assisted by X-ray fluoroscopy and 159 cervical pedicle screws were placed assisted by CT-based navigation system. Screw positions were evaluated by postoperative CT scans or C-arm X-ray 3 dimensional reconstruction. In the computer- assisted group, 155 (97.5%) screws were found completely within their pedicles compared with 133 (91.7%) correctly placed screws in the X-ray assisted group (P < 0.05). The process of navigation investigated in twenty patients showed that the mean time for registration and surface matching was 3.5 (range from 2 to 8 minutes) minutes and the mean time for screw-marker insertion was 2 (range from 1 to 3.5 minutes) minutes with the mean position deviation 0.31 mm (0.12-0.56 mm) per screw. Only twice C-arm fluoroscopy were performed to verify the accuracy of one screw position intraoperatively. No severe complications like neurovascular lesion occurred in both groups. CT-based navigation system can increase the accuracy of cervical pedicle screw implantation significantly.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.