Abstract

The aim of this study was to evaluate the accuracy of the new measurement software, created for aortic stent-graft planning. The software finds semiautomatically the centerline of the vessel from computer tomographic angiography data and calculates semiautomatically diameters and lengths. Two phantoms were constructed, scanned by CT and measured with the measurement software. Results were compared with the known dimensions of the phantoms and mean fractional errors were calculated. Linear regression analysis was also made between measured and known diameters. The mean fractional error (+/- SD) for all diameter measurements combined was 0.017 +/- 0.011. In the linear regression analysis, was 1.000. For length measurements, the mean fractional error (+/- SD) was 0.009 +/- 0.0006. The software enables accurate measurements perpendicular or parallel to the semiautomatically created centerline path.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.