Abstract

Objective: To determine the accuracy of computer-assisted fluoronavigation for percutaneous iliosacral screw placement.Materials and Methods: A fluoronavigation system (Stryker Navigation System, Stryker Leibinger, Kalamazoo, MI) was used to guide the placement of four iliosacral screws into the S-1 bodies of each of five cadaveric pelvic specimens with intact soft tissues. Accuracy of screw placement was verified by radiographs, CT scans, and direct dissection.Results: All 20 screws were placed accurately without complications. Nineteen screws were completely contained within the osseous “safe zone.” On direct dissection, one screw was noted to have penetrated the S-1 foramina by 3 mm without impingement on the nerve root. This was not detected on radiograph or CT scan.Conclusion: The results of this study support the safety and accuracy of computer-assisted fluoronavigation for iliosacral screw placement. The advantages include decreased fluoroscopic time, real-time simultaneous visualization of all three views (inlet, outlet, and lateral), and increased accuracy of placement. Clinical study is warranted.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.