Abstract
Automated blood pressure recorders are used with increasing frequency by pregnant women, mostly without proper evaluation of their accuracy. We compared blood pressures (BP) recorded by 2 automated noninvasive devices, the Spacelabs 90207 ambulatory blood pressure monitor and the OMRON HEM 705 CP portable self-initiated device, with blood pressures recorded by routine sphygmomanometry in 79 pregnant women either considered 'at risk' for preeclampsia or with mild hypertension in pregnancy. The Spacelabs device tended to overestimate systolic BP by a mean 11 (SD=8) mmHg and diastolic BP by 5 (SD=7) mmHg for phase 5 pressure (p<0.001) but was similar to routine BPs for diastolic phase 4 pressures. The OMRON device tended to underestimate diastolic (phase 4) pressure by 4 (SD=6) mmHg (p<0.001) but gave similar systolic and diastolic (phase 5) pressures to routine sphygmomanometry. However, for both devices there was considerable individual patient variability in accuracy. When using these devices to record a limited number of blood pressure recordings, as in this study, we suggest that individual comparison with mercury sphygmomanometry be made in each pregnant woman before accepting the validity of these recordings.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: The Australian & New Zealand journal of obstetrics & gynaecology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.