Abstract

Difficulties are involved in impression making with conventional open impression trays. The purpose of this study was to evaluate the accuracy of transferring implant impressions with a self-perforating impression tray. A reference model of a mandible was fabricated, and 4 implants were placed in the regions of the first premolars and lateral incisors (implants 1, 2, 3, 4). Ten impressions of the reference model with polyvinyl siloxane were made for each group; control (conventional open impression tray) and test (self-perforating impression tray; Miratray Implant). A metal bar was screw-retained on implant 1, and the gaps generated at the vestibular face of implants 3 and 4 were measured by optical microcopy. The 2-way ANOVA and least square difference post hoc test were used (α=.05). Higher mean (±SD) values were obtained for the test group than for the control group for both implants: implant 3: 150 ±84 μm for the test group, 73 ±63 μm for the control group (P=.019); implant 4: 129 ±65 μm for the test group, 62 ±61 μm for the control group (P=.04). The self-perforating impression tray provided less accuracy than the conventional open tray.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.