Abstract
We analyzed the behavior of a single-electron pump operated by sinusoidal signals. We determined the expressions of a set of the harmonic gate signals which can operate an N-junction pump. Concerning the three-junction pump and four-junction pump, we analytically derived asymptotic formulas of the error rate against the pumped current caused by cotunneling and missed transition. As a result, the error rate of the three-junction pump operated by harmonic signals was the same as or less than that of the three-junction pump operated by well-known operation methods, namely, the operation by triangular signals or orthogonal sinusoidal signals. Our estimation shows that the error rate of the three-junction pump is sufficiently small to calibrate the 1.000 00 pA range of a commercially available high-sensitivity current meter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.