Abstract

Objective: The aim of this study was to evaluate the accuracy of a novel 3-dimensional (3D) fluoroscopic navigation system using a flat-panel detector-equipped C-arm, focusing on the influence of the distance from the center of fluoroscopic imaging on navigation accuracy.Materials and Methods: A geometric phantom was made using a Styrofoam cube with 25 markers, each consisting of a metal ball 1.5 mm in diameter, fixed in a cross arrangement at 1-cm intervals. Hip joint surgery was simulated using a set of dry pelvic and femoral bones. A total of eight markers were fixed to the acetabulum and proximal femur.Results: In the geometric phantom study, mean target registration error (TRE) was 0.7 mm (range: 0.1–1.5). The TRE of markers located at 5 cm from the imaging center was significantly higher than the TRE of markers located at 1 and 2 cm. However, the TRE was <1 mm in 90% of the overall trials and <1.5 mm in 100%. In the dry bone study, the mean TRE was 0.9 mm (range: 0.7–1.5) over the acetabulum and 1.0 mm (range: 0.5–1.4) over the femur. No significant difference in TRE was seen between the acetabulum and proximal femur.Conclusion: The accuracy of this novel 3D fluoroscopic navigation system was considered acceptable for clinical application. A 3D C-arm equipped with a flat-panel detector could increase the feasibility of 3D fluoroscopic navigation by reducing the effects of image distortion on navigation accuracy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.