Abstract

IntroductionThe benefits of portable dental X-ray devices remain controversially debated. This study aimed to compare the accuracy in positioning dental X-ray images using handheld (Nomad Pro 2) and wall-mounted (Heliodent Plus) X-ray devices. Materials and methodsRadiographical imaging was exercised on a maxillofacial phantom using the handheld dental X-ray device Nomad Pro 2 (Kavo Kerr, Biberach, Germany) and the wall-mounted dental X-ray device Heliodent Plus (Sirona Dental Systems, Bensheim, Germany). Accuracy of device positioning (i.e., centeredness and perpendicularity) was measured as horizontal and vertical deviation (pixels and millimeters) from a centrally positioned crosshair. The reproducibility of the results was tested for 80 images per device and operator. IBM SPSS (SPSS Inc., Chicago, IL) was used for statistical analyses. ResultsCenteredness of produced dental X-rays was significantly better for the wall-mounted device than for the handheld device in both deviation from the x-axis (p = 0.042) and y-axis (p = 0.020). The perpendicularity of the produced dental X-rays was significantly better for the handheld device than for the wall-mounted device for both horizontal (p < 0.001) and vertical (p < 0.001) plains. ConclusionsHandheld dental X-ray devices appear to provide a high degree of accuracy in image positioning, especially in regard to proper perpendicular image angulation.