Abstract
In this paper, accuracy or error in permeability measurement using a combined microstrip line-coaxial conductor method was investigated. The measurement circuit used in this study is composed of a microstrip line (MSL) circuit and a grounded metal pipe with a center conductor passing through it coaxially. A sample is placed between the metal pipe and the center conductor. We evaluated the measurement accuracy for this measurement arrangement with electromagnetic simulation for the case where there are gaps between the sample and the holder. As a result, it has been shown that the normalized errors for this method have similar gap size dependence to the conventional coaxial method, but are about 10 to 20 percent greater than the coaxial method. With a view to improving the measurement accuracy, a correction method for the error is also discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.