Abstract

The aim of this study is to compare the accuracy of acetabular cup positioning between the accelerometer-based navigation system and the augmented reality-based navigation system during THA in the supine position. This retrospective study included 66 patients (70 hips) who underwent THA using two types of portable navigation system, Hip Align or AR-Hip, in the spine position. The absolute difference between the intraoperative navigation record and postoperative measurement using computed tomography data was evaluated. Preoperative clinical factors that decreased the accuracy of cup positioning by ≥ 3° were analyzed via multiple logistic regression analyses. The average absolute error of inclination was 2.8 ± 2.6° in Hip Align and 2.7 ± 1.8° in AR-Hip, and absolute anteversion error was 2.5 ± 2.0° in Hip Align and 2.6 ± 2.2° in AR-Hip, and there was no significantly different between the two navigation systems. There was a significant association between the absolute measurement error (≥ 3°) of cup inclination and patients' BMI in the Hip Align group [odds ratio (OR) 1.350; 95% confidence interval (CI) 1.035-1.760; p = 0.027], but not in the AR-Hip group. The accuracy of the acetabular cup positioning between the Hip Align and AR-Hip showed no difference during THA in the spine position. The high BMI could have negative influence on the accuracy of cup positioning in THA using Hip Align, thus AR-Hip could be designable for obesity patients.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.