Abstract

Using an iodine-stabilized He-Ne laser as a transfer oscillator, we compare absolute measurements of the optical frequency from a traditional frequency synthesis chain based on harmonic generation and from the frequency division technique of an ultrawide bandwidth femtosecond frequency comb. The agreement between these two measurements, both linked to the Cs standard, is 220+/-770 Hz, yielding a measurement accuracy of 1.6x10(-12). We report 473 612 353 604.8+/-1.2 kHz as a preliminary updated value of the absolute frequency of the " f" component for the He-Ne laser international standard at 633 nm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.