Abstract

The aim of the present study was to measure the accuracy and reproducibility of probe forces in simulated assessments of periodontal pocket depth. The study included experienced and inexperienced examiners and used manual and pressure-sensitive probes. Sixty-one participants were divided into seven groups and asked to probe selected anterior and posterior sites with three different probes (Williams 14W, Chapple UB-CF-15, and Vivacare TPS probes). The model was positioned on a digital electronic balance to measure force, which was recorded initially and after 15min. Probe preferences were recorded. Accuracy was measured by comparing to a standardized 25g force, and reproducibility was calculated for all duplicate measurements. The Vivacare probe produced the most accurate and most reproducible forces, whereas the Williams probe produced the least accurate and least reproducible forces. Probe forces were lighter at anterior sites compared to posterior sites at baseline. Probe forces were reduced at both sites after 15min compared to baseline. Vivacare TPS periodontal probes are more accurate and reproducible than Chapple and Williams probes. Many clinicians in this study preferred the Chapple probe.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.