Abstract

Purpose: The aim of this study was to assess the accuracy of ultra-low dose (ULD) CBCT in orthodontic landmark identification compared with standard dose CBCT on reconstructed lateral and posteroanterior views. Materials and Methods: 10 dry human skulls were scanned twice; once with standard CBCT standard radiation dose, and with ULD CBCT protocol operated at 90 kVp, 7.1 mA, 4.5 sec and 400µm voxel size. From each scan, a lateral and a posteroanterior cephalogram were reconstructed, and 33 landmarks were identified on them by 2 experts. All measurements were then repeated after 2 weeks for intra-observer reliability. Results: There was no statistically significant difference in landmarks identification between ULD and standard dose CBCT. The mean difference between both groups ranged between -0.12 to 0.51 in lateral view and -0.67 to 0.46 in posteroanterior view with excellent inter and intra observer agreement with an ICC value >0.7. Conclusion: Ultra-low dose CBCT with reconstructed lateral and posteroanterior views may be recommended for regular orthodontic diagnostic measurements and can replace the standard dose CBCT

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.