Abstract

The aim of the study was to compare accuracy and reproducibility of four point-of-care testing (POCT) devices (GEM® Premier 3000, ABL 800 flex, GEM® OPL™, HemoCue® B-Hemoglobin) for hemoglobin (Hb) analyzes as compared with the reference laboratory method (Sysmex XE 2100) in children undergoing major surgery. Pediatric patients undergoing craniofacial, spine, hip, or cancer surgery were included. Blood samples for Hb testing were taken at several intraoperative time points and generally withdrawn from the arterial catheter, if accessible. A total of 256 blood samples were taken intraoperatively from 71 pediatric patients. All POCT devices showed very small bias (maximum -0.46 g/dL) to reference method as well as very good reproducibility (maximum coefficient of variation of 0.99%). However, in two cases (HemoCue), potential clinical relevant differences were observed beyond a range of 2 g/dL. All POCT devices tested and operated by trained staff for hemoglobinometry showed reliable test results. They all allow for simple, fast, and precise bedside determination of hemoglobin concentration in the intraoperative setting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.