Abstract

Accurate measurements of in-vivo hip kinematics may elucidate the mechanisms responsible for impaired function and chondrolabral damage in hips with femoroacetabular impingement (FAI). The objectives of this study were to quantify the accuracy and demonstrate the feasibility of using dual fluoroscopy to measure in-vivo hip kinematics during clinical exams used in the assessment of FAI. Steel beads were implanted into the pelvis and femur of two cadavers. Specimens were imaged under dual fluoroscopy during the impingement exam, FABER test, and rotational profile. Bead locations measured with model-based tracking were compared with those measured using dynamic radiostereometric analysis. Error was quantified by bias and precision, defined as the average and standard deviation of the differences between tracking methods, respectively. A normal male volunteer was also imaged during clinical exams. Bias and precision along a single axis did not exceed 0.17 and 0.21 mm, respectively. Comparing kinematics, positional error was less than 0.48 mm and rotational error was less than 0.58°. For the volunteer, kinematics were reported as joint angles and bone-bone distance. These results demonstrate that dual fluoroscopy and model-based tracking can accurately measure hip kinematics in living subjects during clinical exams of the hip.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.