Abstract

To investigate the accuracy of deriving dose-volume histogram (DVH) parameters from deformably registered data by comparing values with the simple addition of DVHs from each phase of a combined external beam radiotherapy (EBRT)/high-dose-rate (HDR-BT) brachytherapy prostate treatment. Eighty-two patients received EBRT in 23 fractions of 2Gy and HDR-BT TG43 in three fractions of 6.5Gy. The HDR-BT CT was deformably registered to the EBRT CT. The rectum D0.1cc , D1cc , D2cc and D10cc were calculated in two ways. (i) Parameter-adding: the EBRT DVH parameters (or the EBRT prescription dose) were added to the unregistered HDR-BT DVH parameters. (ii) Distribution-adding: the parameters were extracted after the EBRT doses were 3D-summed with the registered HDR-BT doses. Resulting differences between the parameters were investigated. The D0.1cc , D1cc and D2cc from parameter-adding were 21.3% (P<0.001), 6.3% (P<0.001) and 3.5% (P<0.001) smaller than those from distribution-adding. The D10cc was 2.2% (P=0.015) larger for distribution-adding. Distribution-adding was confounded by unsystematic inter/intra-observer rectum-contouring errors and registration accuracy near the anterior rectal wall. Consequently, clinical use of distribution-adding to assess rectal doses requires careful contour and registration evaluation.