Abstract

A 3-year-old child was anesthetized for ENT examination and surgery. After induction and tracheal intubation, the patient was ventilated (controlled mode). The respirator screen showed information compatible with a failure of intubation: no expired CO2, no expired flow, no alarm of high pressure limit, and no respiratory chest movement. A fall of SpO2 appeared rapidly which recovered after extubation and manual ventilation through a face mask and reintubation. The expiratory CO2 was present when the patient was ventilated manually and disappeared under controlled ventilation. The increase in the value of the maximal insufflation pressure allowed efficient ventilation with an expiratory CO2 curve and showed high ventilation pressure compatible with a bronchospasme. This case report shows that in case of bronchospasme, if the value of the maximal insufflation pressure is low, this may lead to an erroneous diagnosis of failure of intubation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.