Abstract

ObjectivesTo evaluate whether iodine quantification with Dual-Energy Computed Tomography (DECT) improves the differentiation of accessory spleens (AS) from lymph nodes (LN) compared to CT number measurements. MethodsAbdominal DECT images of 75 patients with either AS (n=35) or LN (n=48) (benign entity) were retrospectively evaluated. Hounsfield Units (HU) and iodine concentrations of AS, LN and the main spleen were measured. Receiver operating characteristics (ROC) were performed to calculate an optimal threshold for distinguishing AS from LN. Sensitivity, specificity, and accuracy were calculated for distinguishing AS from LN by iodine concentration measurements. ResultsMean CT numbers and iodine concentrations were higher for AS (148±29 HU and 48.2±11×100μg/cc) than LN (83±19 HU and 31.5±6.2×100μg/cc, respectively, P<0.001 each). Mean CT numbers were lower for AS compared to the main spleen (161±29HU, P<0.01), whereas mean iodine concentrations (47.7±10×100μg/cc) were not significantly different (P=0.095). An iodine concentration greater than 38×100μg/cc suggested AS with a sensitivity, specificity and accuracy of 91%, 85%, and 88%, respectively (Area under ROC curve 0.941). ConclusionsIodine measurements might contribute to the differentiation of AS from LN. Iodine concentrations similar to that of the main spleen may help to confirm the diagnosis of AS.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.