Abstract

In this study, access flow in hemodialysis patients with bridge grafts was measured by a newly developed device. The technique is based on the measurement of changes in the ultrasound characteristics of blood. These changes were initiated by an injection of isotonic saline and measured in the tubes of the extra-corporeal circuit. The access flow rate (mean +/- standard variation) was 880 +/- 440 mL/min (range, 166 to 1740) (N = 46). The mean coefficient of variation was 13.4 +/- 6.8% (median, 13.0%; range 3.5 to 29.4%). Measurements correlated well (r = 0.91, N = 22, P < 0.001) with flow rates determined by magnetic resonance angiography and by a technique based on intra-access flow-pressure curves (r = 0.84, N = 14, P < 0.001). In conclusion, access flow can be measured easily, noninvasively, and reliably by the ultrasound dilution device.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.