Abstract

Exploit accelerometry data for an automatic, reliable, and prompt detection of spontaneous circulation during cardiac arrest, as this is both vital for patient survival and practically challenging. We developed a machine learning algorithm to automatically predict the circulatory state during cardiopulmonary resuscitation from 4-second-long snippets of accelerometry and electrocardiogram (ECG) data from pauses of chest compressions of real-world defibrillator records. The algorithm was trained based on 422 cases from the German Resuscitation Registry, for which ground truth labels were created by a manual annotation of physicians. It uses a kernelized Support Vector Machine classifier based on 49 features, which partially reflect the correlation between accelerometry and electrocardiogram data. Evaluating 50 different test-training data splits, the proposed algorithm exhibits a balanced accuracy of 81.2%, a sensitivity of 80.6%, and a specificity of 81.8%, whereas using only ECG leads to a balanced accuracy of 76.5%, a sensitivity of 80.2%, and a specificity of 72.8%. The first method employing accelerometry for pulse/no-pulse decision yields a significant increase in performance compared to single ECG-signal usage. This shows that accelerometry provides relevant information for pulse/no-pulse decisions. In application, such an algorithm may be used to simplify retrospective annotation for quality management and, moreover, to support clinicians to assess circulatory state during cardiac arrest treatment.