Abstract
A transition to a mechanical source for x-ray reference field is desired to replace high-activity Co-60 and Cs-137 gamma-ray sources for enhancing nuclear security and safety management. The National Metrology Institute of Japan (NMIJ) developed a new photon reference field using an electron beam from a linac with energy less than 1 MeV. The effective energy of the photon field is comparable with Cs-137 gamma-ray. The air kerma rate in the reference field was validated by comparing a commercial cavity ionization chamber to the NMIJ standard reference field. The calibration constants were consistent within 0.8% for every relative extended uncertainty (k = 2).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.