Abstract

To accelerate dynamic imaging of the vocal tract during articulation. Five subjects were imaged by magnetic resonance imaging (MRI) while repeating the word "Partikel" at 90 beats per minute, using both a radio-frequency-spoiled radial gradient echo sequence with golden angle projection rotation and a previously applied real-time Cartesian sequence. The acquired data were reconstructed using a CG-SENSE method and Cartesian reconstruction. The images from both methods were compared by measuring distances between anatomical landmarks that are important for resonance behavior of the vocal tract. Only commonly available hardware and software were used. With the presented radial method a spatial resolution of 1.8 mm at 25 frames per second could be achieved. Overall, the morphometric measurements of the vocal tract showed less deviation for the radial sequence both across repetitions and for all subjects. Fast modifications, such as complete lip closing, could be analyzed with greater fidelity using the high temporal resolution of the radial sequence. The results indicate that higher temporal resolution is a necessity to catch fast morphometric changes in the vocal tract.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.