Abstract

PurposeRadiochromic films are versatile 2D dosimeters with high‐resolution and near tissue equivalence. To assure high precision and accuracy, a time‐consuming calibration process is required. To improve the time efficiency, a novel calibration method utilizing the ratio of the same dose profile measured at different monitor units (MUs) is introduced and tested in a proton and photon beam.MethodsThe calibration procedure employs the dose ratio of film measurements of the same relative profile for different absolute dose values. Hence, the ratio of the dose is constant at any point of the profile, but the ratio of the net optical densities is not constant. The key idea of the method is to optimize the calibration function until the ratio of the calculated doses is constant. The proposed method was tested in the dose range between 0.25–12 and 1–6 Gy in a proton and photon beam, respectively. A radial symmetric profile and a rectangular profile were created, both having a central plateau region of about 3 cm diameter and a dose falloff of about 1.5 cm at larger distances. The dose falloff region was used as input for the optimization method and the central plateau region served as dose reference points. Only the plateau region of the highest dose entered the optimization as an additional objective. The measured data were randomly split into differently sized training and test sets. The optimization was repeated 1000 times with random start value initialization using the same start values for the standard and the gradient method. Finally, a proton plan with four dose levels was created, which were separated spatially, to test the possibility of a full calibration within a single measurement.ResultsParameter estimation was possible with as low as one dose ratio used for optimization in both the photon and the proton case, yet exhibiting a high sensitivity on the dose level. The root mean squared deviation (RMSD) of the dose was less than 1% when the dose ratio was in the order of 20, whereas the median RMSD of all optimizations was 1.7%. Using four dose levels for optimization resulted in a median RMSD of 1% when randomly selecting the dose levels. Having at least one dose ratio of about 20 included in the optimization considerably improved the RMSD of the calibration function. Using six or eight dose levels reduced the sensitivity on the dose level selection and the median RMSD was 0.8%. A full calibration was possible in a single measurement having four dose levels in one plan but spatially separated.ConclusionsThe number of measurements required to obtain an EBT3 film calibration function could be reduced using the proposed dose ratio method while maintaining the same accuracy as with the standard method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.