Abstract

Rapid identification of microbial pathogens “directly” from positive blood cultures (PBCs) is critical for prompt initiation of empirical antibiotic therapy and clinical outcomes. Towards higher microbial identification rates, we modified a published initial serum separator tubes-based MALDI-TOF-MS protocol, for blood culture specimens received at a non-hospital based standalone diagnostic laboratory, Bangalore, India: (a) “Initial” protocol #1: From 28 PBCs, identification= 39% (Gram-negative= 43%: Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa; Gram-positive: 36%: Enterococcus faecalis, Staphylococcus aureus, Staphylococcus haemolyticus); mis-identification= 14%; non-identification= 47%. (b) “Modified” protocol #2: Quality controls (ATCC colonies spiked in negative blood cultures) From 7 analysis, identification= 100% (Escherichia coli, Klebsiella pneumonia, Klebsiella oxytoca, Pseudomonas aeruginosa, Enterococcus faecalis, Staphylococcus aureus); From 7 PBCs, identification= 57%; mis-identification= 14%; non-identification= 29%. Microbial preparations of highest quality and quantity for proteomic analysis and separate spectra matching reference databases for colonies and PBCs are needed for best clinical utility.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call