Abstract

We performed AC magnetic susceptibility measurement for high frequency (f) range, based on a superconducting quantum interference device (SQUID) magnetometer, MPMS (Quantum Design Inc.). Original system of MPMS uses an A/D converter with 12 bit or 16 bit, and there the frequency of AC field is limited to less than 1 kHz by a software. In the present system, we transformed an analog signal of SQUID output to a digital one via an A/D converter with 24 bit, and generated AC field by a signal generator outside MPMS. The maximum f was enlarged from 1 kHz to 10 kHz. However, it has been confirmed that amplitude of AC field (HAC), at which the rf-SQUID works as a current/voltage transformer, decreases with increasing f. The value of HAC was restricted below 0.02 Oe at f = 10 kHz. The performance of the present system has been confirmed via the measurements of a paramagnetic material Dy2O3 and a rare-earth ferromagnet Ho. The normalization constant for the magnitude of the SQUID output, to calibrate the effect of eddy current, is presented as a function of frequency.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.