Abstract

An AC shunt calibration system using a current‐bridge method is developed, and its measurement uncertainties are estimated. A comparison of the calibration results obtained using the current‐bridge method with the results using DC resistance and AC–DC difference methods was performed to verify the system. These comparison results show that the current‐bridge‐based system performs well with small measurement uncertainties for both the AC resistance and phase angle of the AC shunts up to 1 kHz. © 2014 Institute of Electrical Engineers of Japan. Published by John Wiley & Sons, Inc.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.