There are 5 known Korotkoff sounds which occur during a standard blood pressure (BP) measurement cycle, the first 4 (K1 to K4) having specific acoustic signatures, and the fifth (K5) occurring where sound first becomes absent. The different Korotkoff sound proportions (KSPs) are not well understood. Using bespoke acoustic analysis technology, we compared KSPs (K1P, K2P, K3P, K4P) in normotensive men and women at different ages and described the effect of normalising the cuff deflation rate (CDR) to allow 60 heart beats (CDR 60 ) versus a standard 3mmHg/s (CDR 3mmHg ). At a CDR 3mmHg , men had a higher K2P (36% vs 26%, p=0.006) and lower K3P (27% vs 38%, p=0.002) than women. K3P significantly increased with age in both men (p<0.001) and women (p<0.004). CDR 60 caused a significant drop in K2P (p<0.001) and rise in K3P (p=0.019). These differences may reflect changes in vascular stiffness which may vary with gender, age and CDR. To our knowledge, this phenomenon has never been described before and requires further investigation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.