Abstract
Short liquid core waveguides (LCWs) were included into a fiber-loop cavity ring-down absorption spectrometer to reduce the detection limit over, both, single pass absorption in a LCW and cavity-enhanced absorption using a conventional fiber-loop cavity. LCWs of 5 and 10 cm length were interfaced with a pressure-flow system and a multimode fiber-loop cavity using concave fiber lenses with matching numerical apertures and diameters. Two red dyes, Allura Red AC and Congo Red, were detected with a 532 nm pulsed laser at a 5 nM limit of detection in a detection volume of less than 1 μL, corresponding to a minimal detectable absorbance of less than 4 × 10(-4) cm(-1) and a minimal detectable change in absorption cross section, σ(min) = V(det) × ε × C(LOD), of about 14 μm(2) (Allura Red AC) and 37 μm(2) (Congo Red).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.