Abstract

The technique presented is an approach to absorp tion edge transmission imaging which eliminates the need for monoenergetic x-ray sources, detector scanning, and energy analysis of transmitted x-rays. Subtraction of images formed with heavily filtered x-ray beams peaked at energies above and below the K absorption edge is accomplished with image-intensified fluoroscopy equipment coupled to a pair of storage tubes. Images of iodine and xenon concentrations of 1 mg/cm2 have been imaged in 15 cm phantoms in a few seconds. When automated, the system will provide similar sensitivity in 20 cm phantoms, with imaging times of 1 second or less with resolution on the order of 1 mm. Several preliminary images are presented which relate to possible future clinical studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.