Abstract
Subtraction and ratioing of strong absorption bands in Fourier transform infrared (FT-IR) spectroscopy produces anomalous absorption errors. One source of error is the instability in the wavenumber scale of the FT-IR spectra. The possible causes of this error are explored. The thermal expansion and contraction of the cavity of the HeNe reference laser from a typical commercial instrument was found to produce changes in the laser wavenumber of ±0.034 cm(-)(1). Changes of this size are shown to introduce errors into the wavenumber scales of FT-IR spectra which are sufficient to produce the observed anomalies. The dependence of the error on instrumental and spectroscopic parameters is explored. Solutions to the problem are proposed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.