Abstract

The measured and theoretically predicted values of detective quantum efficiency (DQE) for a CsI x-ray image intensifier are compared for nine monoenergetic beams of x rays. The agreement between measurement and theory of better than +/- 5% indicates that we have a sound understanding of the physical parameters controlling the DQE. It is shown that the fraction of K-fluorescent x rays escaping from the input phosphor is independent of incident energy. The number of electrons released within the x-ray image intensifier (XRII) by an incident x ray has been measured. The mechanism for energy broadening within the XRII is shown to be predominantly the limited number of electrons and not light absorption.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call