Abstract

Mass absorption coefficient data published by Allen, Bragg, and Victoreen for the wavelength region 0.1 to 50 A are plotted for several common x-ray tube and counter windows (Al, Be, and mica) and gases (Xe, Kr, A, and Ne) used in x-ray detectors. A graph of the calculated percentage transmission of the windows is presented. The calculated quantum-counting efficiencies for the range 0.1 to 15 A are shown for an end-window argon-filled Geiger counter, side-window Xe, Kr, A, and Ne proportional counters and an NaI·Tl scintillation counter. Some of the factors to be considered in the choice of detectors for x-ray diffractometry and x-ray spectroscopy are discussed particularly with respect to the spectral sensitivity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.