Abstract

Irradiation of blood bags using X-ray irradiators and dosimetry services are required to ensure uniform dose levels in the range 25–50 Gy to prevent Transfusion Associated Graft versus Host Disease (TA-GvHD). An absorbed dose characterization of a Raycell MK2 X-Irradiator was performed using three different dosimetric systems. Results showed a dosimetric accuracy of the ionization chamber together with the Alanine dosimeter. TLDs measurements exhibited a small overestimation by 4% of the absorbed dose. The Dose Uniformity Ratio (DUR), between maximum and minimum dose levels in the canister, was in good agreement with the manufacturer specifications (≤1.5).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.