Abstract

Volumetric CT using a cone beam has been developed by several manufacturers for dentomaxillofacial imaging. The purpose of this study was to measure doses for implant planning with cone beam volumetric imaging (CBVI) in comparison with conventional multidetector CT (MDCT). The two CBVI systems used were a 3D Accuitomo (J. Morita), including an image-intensifier type (II) and a flat-panel type (FPD), and a CB MercuRay (Hitachi). The 3D Accuitomo operated at 80 kV, 5 mA and 18 s. The CB MercuRay operated at 120 kV, 15 mA, 9.8 s. The MDCT used was a HiSpeed QX/i (GE), operated at 120 kV, 100 mA and 0.7 s, and its scan length was 77 mm for both jaws. Measurement of the absorbed tissue and organ doses was performed with an Alderson phantom, embedding the radiophotoluminescence glass dosemeter into the organs/tissues. The values obtained were converted into the absorbed dose. The effective dose as defined by the International Commission on Radiological Protection was then calculated. The absorbed doses of the 3D Accuitomo of the organs in the primary beam ranged from 1-5 mGy, and were several to ten times lower than other doses. The effective dose of the 3D Accuitomo ranged from 18 muSv to 66 muSv, and was an order of magnitude smaller than the others. In conclusion, these results show that the dose in the 3D Accuitomo is lower than the CB MercuRay and much less than MDCT.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.