Abstract

The absolute standardization of Co-57 by the 4π(PC)-γ coincidence method, in efficiency extrapolation variant, is presented. An original, improved method, based on the use of the equivalent decay scheme, was applied; apart from the enhancement of efficiency, a diminution of extrapolation slope was obtained, lowering the extrapolation uncertainty. The results obtained in the participation at the supplementary comparison CCRI(II)-S6.Co-57 and at the key comparison code BIPM.RI(II)-K1.Co-57 are presented. The method allows to simplify the treatment of 56Co and 58Co impurities’ correction, which influences significantly the measurements in a CENTRONIC type ionization chamber. An example of treatment of this mixture, where the influence of impurities on results was analyzed, will be presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.