Abstract

The absolute standardization of Co-57 by the 4π(PC)-γ coincidence method, in efficiency extrapolation variant, is presented. An original, improved method, based on the use of the equivalent decay scheme, was applied; apart from the enhancement of efficiency, a diminution of extrapolation slope was obtained, lowering the extrapolation uncertainty. The results obtained in the participation at the supplementary comparison CCRI(II)-S6.Co-57 and at the key comparison code BIPM.RI(II)-K1.Co-57 are presented. The method allows to simplify the treatment of 56Co and 58Co impurities’ correction, which influences significantly the measurements in a CENTRONIC type ionization chamber. An example of treatment of this mixture, where the influence of impurities on results was analyzed, will be presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.