Abstract
We have developed a new cryogenic radiometer for determining absolute power. This instrument is composed of a superconducting transition edge sensor as a thermometer, a silicon substrate as a heat absorber, and a SQUID amplifier as a readout. The input power is measured with electrical substitution of a heater on the device. By using very sharp transition of the thermometer, the device temperature is locked within its transition temperature. This device demonstrated very wide dynamic range of more than 105, very low noise equivalent power of 1.5 nW. These properties are very attractive to be used as highly precise power measurement instruments in metrological fields.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.