Abstract
An apparatus for precise and absolute measurement of the thermal conductivity of electrically conducting liquids using the transient hot-wire method has been developed. In the present apparatus, a metallic wire coated with a thin electrical insulation layer has been used as a heating element and a resistance thermometer instead of a bare metallic wire. The effects on the thermal conductivity measurement caused by the thin insulation layer have been analysed. In the analysis, it was found that the effects can be negligibly small if the instrument is adequately designed. The usability of the method for electrically conducting liquids has been tested to measure the thermal conductivity of an aqueous NaCl solution in the temperature range 0 to 45 degrees C at atmospheric pressure. The accuracy of the present measurement was estimated to be +or-0.5%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.