Abstract

A measuring set-up comprising a proportional counter of calculable 222 Rn efficiency and quantifiable active volume ( δ V<0.1%) is described. On account of the special design of the end caps, the counter is suitable for absolute activity measurements on gaseous radiation sources. The 222 Rn efficiency is determined by computer simulation of the measured α-spectra. The procedures necessary for absolute measurements by means of the counter are described, and the suitability of the counter for absolute measurements of the 222 Rn activity is proved by experiments. Thus, a new method for the realization of the unit of activity of 222 Rn is obtained, which is independent of the unit of activity of 226 Ra .

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call