Abstract
Gravimetric synthetic mixtures prepared from highly enriched isotopes of zinc in the form of oxides of well defined purity were used to calibrate a thermal ionization mass spectrometer. Measurements on natural zinc samples yielded an absolute isotopic composition of 48.27 (21) at. % 64Zn, 27.98 (5) at. % 66Zn, 4.10 (1) at. % 67Zn, 19.02 (8) at. % 68Zn, and 0.63 (1) at. % 70Zn, and the atomic weight of zinc as 65.409 (6) both with an uncertainty given on the basis of 95% confidence limit.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.