Abstract
The paper presents the results of absolute spectral responsivity measurements for SPPD11 and SPPD11-04 pulse-type silicon detectors, carried out in the X-ray energy range from 0.3 to 1.5 keV with a relative uncertainty of about 8% (1 σ). The measurements were performed using the detector exposure to a well-calculable synchrotron radiation through the selective filters with well-known transmittance. Synchrotron radiation from the VEPP-2M storage ring was utilized for these measurements. The spectral responsivity of the detectors was restored by solving the proper system of integral equations on the base of measurement data. For increasing the calibration accuracy, the X-ray transmittance near the L-absorption edges of the filters used and the angular spread in the positron beam of the storage ring were experimentally determined.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.