Abstract

In a Fizeau interferometer measurement, one of the factors that reduces long-term measurement reproducibility is variation in the shape of the reference flat. The purpose of the present work is to improve the measurement accuracy of a Fizeau interferometer by establishing the causes of the reference flat deformation. We discovered that the deformation is caused by a change in environmental humidity. We confirmed that a reference flat with a diameter of 152mm deforms to a paraboloidal shape with a peak-to-valley height of 1nm in response to a relative humidity change of 11%. The cause of the deformation can be interpreted as being a change in the film stress in an antireflective coating on the reverse side of the reference flat due to the adsorption of water molecules. We carried out the three-flat test for an absolute calibration of optical flats by considering the effect of a relative humidity change. The calibration results indicate that the uncertainty in the measurements of the Fizeau interferometer was successfully improved to 0.7nm (σ: standard deviation) along a length of more than 145mm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.