Abstract

A proton energy spectrometer system is composed of a time-of-flight spectrometer (TOFS) and a Thomson parabola spectrometer (TPS), and is used to characterize laser-accelerated protons. The TOFS detects protons with a plastic scintillator, and the TPS with a CR-39 or imaging plate (IP). The two spectrometers can operate simultaneously and give separate time-of-flight (TOF) and Thomson parabola (TP) data. We propose a method to calibrate the TOFS and IP by comparing the TOF data and the TP data taken with CR-39 and IP. The absolute response of the TOFS as a function of proton energy is calculated from the proton number distribution measured with CR-39. The sensitivity of IP to protons is obtained from the proton number distribution estimated with the calibrated TOFS. This method, based on the comparison of the simultaneously measured data, gives more reliable results when using laser-accelerated protons as a calibration source. The calibrated spectrometer system can be used to measure absolutely calibrated energy spectra for the optimization of laser-accelerated protons.