Abstract

Absolute calibration of a soft X-ray spectrograph has been performed using a white beam of synchrotron radiation. The calibrated spectrograph was a flat-field grazing-incidence spectrograph with an X-ray CCD detector for X-ray laser research. Absolute sensitivity of the spectrograph system can be obtained from transmitted spectra using filters made of several different materials, each providing an absorption-edge wavelength standard. The absolute sensitivity determined in this work shows nearly the same behaviour with wavelength as that in another calibration experiment using a laser-produced plasma as an X-ray source.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.