Abstract

The sum-peak method was applied to calibrate four 60Co sources in the range of 25–350 kBq. An HPGe-based γ-ray spectroscopy system was used, and the effects of the angular correlation and pulse pile-up were corrected, and are discussed. The correction for the angular correlation was obtained from a theoretical calculation, while the pulse pile-up effect was corrected using a new extrapolation technique based on the pile-up measurements by varying the shaping time of the amplifier. The determined radioactivities were within 2% deviation from the values reported by the supplier and the associated uncertainties were less than 2%. The observed consistency was obtained under the experimental condition of keeping the total counting rate ⩽15 kcps.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.